OPERATING INSTRUCTIONS
30 Sep 13 4-21
DE-7200VFDCentrifuge
NORMAL SHUTDOWN
The normal shutdown procedure is to be used for controlled stopping of operation. Normal
shutdown is performed for routine activities such as cleaning, lubrication, inspection, or
adjustment.
Note!
Proper shutdown and flushing of the bowl can prevent high vibration at
the next startup.
Step Procedure
1 Stop the feed pump.
2
Open supply of fresh flushing liquid to remove all solids from conveyor and bowl.
Continue flushing for 3 to 5 minutes after feed pump has been shut down. Regardless of
shutdown duration, the conveyor operates at a pre-set differential speed while the bowl
decelerates to a stop. This process cleans the bowl during the shutdown.
As the speed gradually drops during the shutdown period, the solids chute can also be
flushed. While bowl is rotating, attach a hose to fitting on the hinge side of the case
(Figure 3-11) to wash the exterior of the rotating assembly and interior of the case
assembly.
3
Select STOP on Operation screen to shut down centrifuge. When bowl speed falls to
less than 200RPM, conveyor gearbox drive motor automatically shuts down.
4
For cabinet with purge system, maintain air supply to cabinet purge unit to keep cabinet
pressurized.
AUTOMATIC SHUTDOWN
The centrifuge has built-in safety features to protect the equipment. These features will result in
automatic shutdown of the centrifuge before damage occurs. The following paragraphs describe
these automatic shutdowns.
Excessive Vibration
Excessive vibration of the centrifuge will cause the vibration switch to interrupt electric power to
the centrifuge run relay, shutting down the machine. Such excessive vibration may occur during
startup or normal operation due to slumping of the wall cake or other unbalanced condition of the
bowl. The machine may be re-started by pressing the reset button on the vibration switch and
then using the Normal Startup procedure above.
If the machine continually trips during normal startup, flush the bowl with fresh water while running
the automatic cleanout routine.
Main Bearing Temperatures
Temperature sensors are installed on the liquid and solid end main bearings and connected to the
PLC. Bearing temperatures are continuously displayed on the Operation screen. An alarm
message is displayed on the control panel when either bearing temperature exceeds 225°F
(107°C). If temperature rises to 250°F (120°C), a fault message appears and the centrifuge is
shut down. Excessively high bearing temperatures usually indicate bearing failure, which can
result from inadequate or excessive lubrication, contamination, or severe wear. Contact Derrick
Service department for assistance.