SECTION 3 - INSTALLATION
01 Jan 14 3-1
DE-7200VFDCentrifuge
GENERAL
This section describes the recommended installation procedure for the DE-7200 VFD centrifuge.
The centrifuge is shipped fully assembled. For safety during shipment, however, the rotating
assembly is supported by shipping brackets. It must be lowered onto the base and then its
bearing pillow blocks secured to the base.
SAFETY
Read and understand ALL safety information presented in this manual before installing and
operating this equipment. Refer to Section 2 for a summary of Warnings addressing installation,
operation, and maintenance of this equipment.
Before beginning the installation, review the equipment handling procedures in this section.
Failure to observe proper equipment handling procedures may result in serious personal injury or
death and/or damage to the equipment.
WARNING! TO ENSURE PROPER BALANCE AND ORIENTATION WHEN UNIT
IS RAISED AND PREVENT DAMAGE TO COMPONENTS, ATTACH LIFTING
SLING ONLY AT DESIGNATED LIFT POINTS. DO NOT ATTEMPT LIFTING BY
ATTACHMENT TO ANY OTHER LOCATION.
WARNING
! BE SURE THAT HANDLING DEVICES HAVE SUFFICIENT LIFTING
CAPACITY TO SAFELY HANDLE THE WEIGHT OF THE EQUIPMENT.
LOWER
THE CENTRIFUGE GENTLY INTO PLACE, AS JARRING MAY CAUSE
DAMAGE.
WARNING! DO NOT REMOVE SHIPPING BRACKETS UNTIL
CENTRIFUGE
HAS BEEN POSITI
ONED AT FINAL INSTALLATION SITE.
WARNING!
WHILE OPENING AND CLOSING THE CENTRIFUGE COVER,
PERSONNEL MUST ENSURE THAT ALL BODY LIMBS AND ANY OTHER
ITEMS ARE CLEAR OF THE COVER GAP. TWO
OPERATORS MUST BE
USED, WITH BOTH HANDS ON THE COVER HANDLES AT ALL T
IMES.
DERRICK RECOMMENDS HAVING A HOIST AVAILABLE I
F ADDITIONAL
ASSISTANCE IS DESIRED DURING THE OPEN
ING/CLOSING PROCEDURE.
Przeglądanie stron 20
1 2 ... 16 17 18 19 20 21 22 23 24 25 26 ... 166 167

Komentarze do niniejszej Instrukcji

Brak uwag